Copyright © All rights reserved. Created By Mirad September 2011
Contec
CMS50A Pulse Oximeter
Major Features
■ Integrated with SpO2 probe and processing display module
■ Small
in volume、light in weight and convenient in carrying
■ Operation of the product is
simple , low power consumption
■ SpO2 value display
■ Pulse rate value display, bar
graph display
■ Pulse waveform display
■ Battery capacity indication
■ With backlight
■
Automatically power off function: when the device is under the state of measuring
interface . it will automatically power
off within 5 seconds if the finger falls
out of probe
Main performance
Display Mode: 1.3" LCD display
Screen Resolution: 128*64
SpO2 Measuring
Range: 0%~100%, (the resolution is 1%).
Accuracy:70%~100%: ±2% , Below 70% unspecified.
PR
Measuring Range: 30bpm~250bpm, (the resolution is 1bpm)
Accuracy: ±2bpm or ±2% (select
larger)
Measurement Performance in Weak Filling Condition:
SpO2 and pulse rate can
be shown correctly when pulse-filling ratio is 0.4%. SpO2 error is ±4%, pulse rate
error is ±2 bpm or ±2% (select larger).
Resistance to Surrounding light: The deviation
between the value measured in the condition of man-made light or indoor natural light
and that of darkroom is less than ±1%.
Power Consumption:Smaller than 40mA.
Voltage:
DC 3.0V
Power Supply: 1.5V (AAA size) alkaline batteries × 2
Battery Working Hour:
The minimum continually work time is 15 hours , theoretical number is 28 hours.
Safety
Type: Interior Battery, BF Type
Accessories
Sell in standard:
■ A hanging rope
■ A user manual
Physical Identity
Dimension: 65(L) × 35(W) × 40 (H) mm
Weight: About 75g (with the
batteries)
Qualification
Passed CE
Plus Oximetrar
Feature